CN | EN
Electrode Wire harness+AdapterCable


*Easy to use:Cable integrated design,easy to plug and unplug

*Comfortable to wear:The lightweight electrode wire,reducing foreign body 

sensation on the head

*Scientific design:according the 10-20 system to customize the each electrode wire length


*Adapter cable:Suitable for all EEG instruments.




EEG (EMG) lead wire

Main structure

Composed of instrument plug, signal wire (shielded and unshielded), metal electrode parts (metal pin, disk, clip, button).

Product performance

1. Applied 10Hz sine wave and not exceed 100μA (p-p) current to electrode pair, The AC impedance of the electrode pair does not exceed 3kΩ.

2. After 1 min stable period, the DC offset voltage of the electrode pair is not more than 30mV.

3. The DC impedance at both ends of the wire is 0.5Ω-2.5Ω (for 1.5 m, when the length of the wire increases, the corresponding resistance increase should not exceed 1Ω/m).

Application

Used with a variety of EEG, EEG Sleep monitors, EEG analyzers, and EMG analysis systems to acquiring and transmit human bioelectric signals.

Contraindications

It is forbidden to use this product if the skin has trauma or has not healed by surgical operation.

Attention and warning

1.Choose the appropriate lead wire according to your needs.

2.Do not use in the presence of flammable anesthetic gas mixed with air or flammable anesthetic gas mixed with oxygen or nitrous oxide.

Instructions

1.Clean the electrode position skin of the patient

2.Connect the lead wires

* Model GCModel SCModel SCCModel SSCModel ABSModel EC

Apply conductive paste to the electrode and fix it at the signal acquisition place, and connect it to the relevant testing instrument through a connecting wire* Model AC

Clamp the product on the electrode to be connected, and connect the other end to the relevant testing instrument 

* Model LModel A

Connect one end of the product to needs to be extended or transferred port, and the other end to the relevant testing instrument 

* Model BE

After soaking the product in nomorl saline, fix it at the signal acquisition place, and connect it to the relevant testing instrument through the connection wire                                                                                                                                   

3.Start recording 

4.After the recording is completed, remove the lead wire and help the patient remove the paste from the skin

5.Clean the electrode with water and dry it for later use.

Product transportation

The product should be protected from squeezing, rainwater, and high temperature heat during transportation.

Product maintenance

The product is cleaned regularly with water. Dry in the shade in a ventilated place. Avoid squeezing.

Storage

Store in a temperature of -10℃+40℃, relative humidity ≤80%, no corrosive gas, and good ventilation.

Validity period3 years

Production DateRefer packaging

Production lotRefer packaging

Company

NameQingdao Tenocom Medical Technology CO.LTD 

AddressRoom 211,Floor 2,No.917 Weihe Road,Huangdao,Qingdao,Shandong Province,China 

Tel0086-532-8695-9323

Manufacturer

NameQingdao Tenocom Medical Technology CO.LTD 

AddressRoom 211,Floor 2,No.917 Weihe Road,Huangdao,Qingdao,Shandong Province,China 

Production addressRoom 209,Floor 2,No.917 Weihe Road,Huangdao,Qingdao,Shandong Province,China

Tel:0086-532-8695-9323

Product serviceQingdao Tenocom Medical Technology CO.LTD